Detection of mycoplasma contamination in cell cultures by a mycoplasma group-specific PCR

Abstract
The suitability of a 16S rRNA-based mycoplasma group-specific PCR for the detection of mycoplasma contamination in cell cultures was investigated. A total of 104 cell cultures were tested by using microbiological culture, DNA fluorochrome staining, DNA-rRNA hybridization, and PCR techniques. A comparison of the results obtained with these techniques revealed agreement for 95 cell cultures. Discrepant results, which were interpreted as false negative or false positive on the basis of a comparison with the results obtained with other methods, were observed with nine cell cultures. The microbiological culture technique produced false-negative results for four cell cultures. The hybridization technique produced false-negative results for two cell cultures, and for one of these cell cultures the DNA staining technique also produced a false-negative result. The PCR may have produced false-positive results for one cell culture. Ambiguous results were obtained with the remaining two cell cultures. Furthermore, the presence of contaminating bacteria interfered with the interpretation of the DNA staining results for 16 cell cultures. For the same reason the hybridization signals of nine cell cultures could not be interpreted. Our results demonstrate the drawbacks of each of the detection methods and the suitability of the PCR for the detection of mycoplasmas in cell cultures.