Patient neutron dose equivalent exposures outside of the proton therapy treatment field
- 20 December 2005
- journal article
- research article
- Published by Oxford University Press (OUP) in Radiation Protection Dosimetry
- Vol. 115 (1-4), 154-158
- https://doi.org/10.1093/rpd/nci264
Abstract
A large fraction of dose to healthy tissue located outside of the treatment field during proton therapy is attributable to neutrons produced in the beam-delivery apparatus. In this work, the neutron dose equivalent (H) per therapeutic proton absorbed dose (D) was estimated for typical treatment conditions as a function of range modulation width, angle with respect to the incident proton beam, and the distance from the isocentre at the Harvard Cyclotron Laboratory's (Cambridge, MA) passively spread treatment field using Monte Carlo simulations. For a beam with 16 cm penetration (depth) and a 5 × 5 cm2 lateral field size at the patient location along the incident beam direction at 100 cm from the isocentre, the predicted H/D values are 0.35 and 0.60 mSv Gy−1 from the simulations and measurements, respectively. At all locations, the predicted H/D values are within a factor of 2 and 3 of the measured result for no modulation and 8.2 cm of modulation, respectively.Keywords
This publication has 12 references indexed in Scilit:
- Design tools for proton therapy nozzles based on the double-scattering foil techniqueRadiation Protection Dosimetry, 2005
- Virtual commissioning of a treatment planning system for proton therapy of ocular cancersRadiation Protection Dosimetry, 2005
- Dosimetry for ocular proton beam therapy at the Harvard Cyclotron Laboratory based on the ICRU Report 59Medical Physics, 2002
- Neutron shielding verification measurements and simulations for a 235-MeV proton therapy centerNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2002
- Measurement of neutron dose equivalent to proton therapy patients outside of the proton radiation fieldNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2001
- A Randomized Controlled Trial of Varying Radiation Doses in the Treatment of Choroidal MelanomaArchives of Ophthalmology (1950), 2000
- Secondary Dose Exposures During 200 MeV Proton TherapyRadiation Protection Dosimetry, 1997
- Definitive radiation therapy for chordoma and chondrosarcoma of base of skull and cervical spineJournal of Neurosurgery, 1982
- Flattening of proton dose distributions for large‐field radiotherapyMedical Physics, 1977
- Radiological Use of Fast ProtonsRadiology, 1946