An experimental investigation of the effect of light-box luminance on the detection of low contrast objects in mammography
- 1 February 1996
- journal article
- Published by Oxford University Press (OUP) in The British Journal of Radiology
- Vol. 69 (818), 153-159
- https://doi.org/10.1259/0007-1285-69-818-153
Abstract
It has been hypothesized that light-box luminance is an important factor in the detection of objects on radiographs. In this work, existing psychophysical data relating to the measurement of visual thresholds at various scene luminance levels are applied to the problem of the observation of radiographs on a light-box. These data suggest that for a given stimulus size, the threshold contrast varies little over several orders of magnitude of scene luminance. In a series of contrast detail experiments performed over a wide range of light-box luminances it has been demonstrated that the detection of low contrast objects on mammographic film is dominated by external noise, that is noise on the film, rather than the internal visual noise of the observer. It is therefore suggested that in mammography it is inappropriate to base recommendations for optimal values of light-box luminance on psychophysical studies of visual noise. It has been shown that commonly used light-box luminances are suitable for viewing mammograms at the higher average optical densities now being recommended, provided that precautions are taken to avoid glare and reflection.Keywords
This publication has 10 references indexed in Scilit:
- The use of a contrast–detail test object in the optimization of optical density in mammographyThe British Journal of Radiology, 1995
- Mammographic film density and detection of small breast cancersClinical Radiology, 1994
- Relationship Between Medical Requirements and Technical Parameters of Good Imaging Performance and Acceptable DoseRadiation Protection Dosimetry, 1993
- Quality assurance of viewing boxes: proposal for establishing minimum requirements and results from a Spanish quality control programmeThe British Journal of Radiology, 1990
- Analysis of variations in contrast‐detail experimentsMedical Physics, 1984
- Quantitative relation between detectability and noise powerPhysics in Medicine & Biology, 1983
- Radiographic mottle and patient exposure in mammography.Radiology, 1982
- Computation of bremsstrahlung X-ray spectra and comparison with spectra measured with a Ge(Li) detectorPhysics in Medicine & Biology, 1979
- Contrast Thresholds of the Human EyeJournal of the Optical Society of America, 1946
- INTENSITY DISCRIMINATION IN THE HUMAN EYEThe Journal of general physiology, 1938