Computation of bremsstrahlung X-ray spectra and comparison with spectra measured with a Ge(Li) detector
- 1 May 1979
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 24 (3), 505-517
- https://doi.org/10.1088/0031-9155/24/3/002
Abstract
A method of computing theoretical X-ray spectra in the range 30-150 kV is presented. The theoretical spectra are compared with constant potential, high resolution spectra from a tungsten target measured with a Ge(Li) detector, for a range of target angles, tube voltage and filtrations. Above 100 kV the spectra were also measured with a NaI detector but, as there was good agreement between the Ge(Li) and NaI detectors, only the former are presented. Spectra computed using Kramers' theory are also included for comparison, giving fairly good agreement at large target angles (30 degrees ) but becoming gradually worse as the target angle decreased. Spectra may be computed by this method for any desired filtration, target angle, and tube voltage between 30 and 150 kV, in excellent agreement with the measured data.Keywords
This publication has 25 references indexed in Scilit:
- A determination by an analysis of X-ray attenuation in aluminium of the intensity distribution at its point of origin in a thick tungsten target of bremsstrahlung excited by constant potentials of 60-140 kVPhysics in Medicine & Biology, 1977
- The spectrum and intensity of extra-focal (off-focus) radiationThe British Journal of Radiology, 1976
- A technique for measuring the variation of photon spectrum from X-ray generators over the mains voltage cyclePhysics in Medicine & Biology, 1976
- Computed spectra from diagnostic and therapeutic X-ray tubesPhysics in Medicine & Biology, 1973
- The effect of target absorption on the attenuation characteristics of bremsstrahlung generated at constant medium potentialsJournal of Physics B: Atomic and Molecular Physics, 1972
- Theoretical continuous and L-characteristic X-ray spectra for tungsten target tubes operated at 10 to 50kVPhysics in Medicine & Biology, 1970
- Photon cross sections from 1 keV to 100 MeV for elements Z=1 to Z=100Atomic Data and Nuclear Data Tables, 1970
- Measurements of RelativeRadiative Decay RatesPhysical Review B, 1969
- The ratio of K characteristic to total radiation emitted from a tungsten target X-ray tubeJournal of Physics D: Applied Physics, 1968
- Transmission of Slow Electrons through Thin FilmsProceedings of the Physical Society, 1960