Accuracy of inhomogeneity correction in photon radiotherapy from CT scans with different settings
- 6 August 2002
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 47 (17), N223-N231
- https://doi.org/10.1088/0031-9155/47/17/402
Abstract
We report an investigation on the accuracy of inhomogeneity correction in photon radiotherapy from CT scans with different settings. Specifically, the dosimetric differences from different CT scan parameters (kV, mAs) to phantoms and from different Hounsfield unit versus electron density (HU-ED) curves to patients are investigated. The absolute dose per monitor units (dose/MU) is used to quantify the results. We found that only for high-density bones (cranium, femoral tube, etc) using small field 18 MV beams, the dose/MU is up to 2% higher for CT scans using 80 kV than for 130 kV at a depth just beyond the bone and is up to 1-1.5% higher for CT scans using 80 mAs than for 300 mAs. For low-density bones (such as femoral head) and lung, the difference is 1% or less with different kV or mAs settings. The dose/MU varies with different HU-ED curves by up to 2%. The HU-ED curve from the stochiometric calibration was found to be more accurate based on a real measurement. A simplified 4-point curve provides nearly the same accuracy as the stochiometric calibration and may be used as an alternative for routine clinical application.Keywords
This publication has 13 references indexed in Scilit:
- The effect of dose calculation accuracy on inverse treatment planningPhysics in Medicine & Biology, 2002
- Comparison of the Batho, ETAR and Monte Carlo dose calculation methods in CT based patient modelsMedical Physics, 2001
- Verification of lung dose in an anthropomorphic phantom calculated by the collapsed cone convolution methodPhysics in Medicine & Biology, 2000
- Photon dose calculation of a three‐dimensional treatment planning system compared to the Monte Carlo code BEAMMedical Physics, 2000
- Feasibility of megavoltage portal CT using an electronic portal imaging device (EPID) and a multi-level scheme algebraic reconstruction technique (MLS-ART)Physics in Medicine & Biology, 1998
- A cone-beam megavoltage CT scanner for treatment verification in conformal radiotherapyRadiotherapy and Oncology, 1998
- The indirect use of CT numbers to establish material properties needed for Monte Carlo calculation of dose distributions in patientsMedical Physics, 1998
- Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous mediaMedical Physics, 1989
- Acceptance testing computerized radiation therapy treatment planning systems: Direct utilization of CT scan dataMedical Physics, 1985
- Compton scatter imaging of transverse sections: an overall appraisal and evaluation for radiotherapy planningPhysics in Medicine & Biology, 1981