Collision‐induced fragmentation pathways including odd‐electron ion formation from desorption electrospray ionisation generated protonated and deprotonated drugs derived from tandem accurate mass spectrometry
- 19 September 2006
- journal article
- research article
- Published by Wiley in Journal of Mass Spectrometry
- Vol. 41 (10), 1277-1286
- https://doi.org/10.1002/jms.1091
Abstract
The rapid desorption electrospray ionisation (DESI) of some small molecules and their fragmentation using a triple-quadrupole and a hybrid quadrupole time-of-flight mass spectrometer (Q-ToF) have been investigated. Various scanning modes have been employed using the triple-quadrupole instrument to elucidate fragmentation pathways for the product ions observed in the collision-induced dissociation (CID) spectra. Together with accurate mass tandem mass spectrometry (MS/MS) measurements performed on the hybrid Q-ToF mass spectrometer, unequivocal product ion identification and fragmentation pathways were determined for deprotonated metoclopramide and protonated aspirin, caffeine and nicotine. Ion structures and fragmentation pathway mechanisms have been proposed and compared with previously published data. The necessity for elevated resolution for the differentiation of isobaric ions are discussed. Copyright © 2006 John Wiley & Sons, Ltd.Keywords
This publication has 17 references indexed in Scilit:
- Ambient Mass SpectrometryScience, 2006
- Desorption electrospray ionization mass spectrometry for the analysis of pharmaceuticals and metabolitesRapid Communications in Mass Spectrometry, 2005
- Rapid accurate mass desorption electrospray ionisation tandem mass spectrometry of pharmaceutical samplesRapid Communications in Mass Spectrometry, 2005
- Desorption Electrospray Ionization Mass Spectrometry for High-Throughput Analysis of Pharmaceutical Samples in the Ambient EnvironmentAnalytical Chemistry, 2005
- Versatile New Ion Source for the Analysis of Materials in Open Air under Ambient ConditionsAnalytical Chemistry, 2005
- A high‐throughput liquid chromatography/tandem mass spectrometry method for screening glutathione conjugates using exact mass neutral loss acquisitionRapid Communications in Mass Spectrometry, 2005
- Direct, trace level detection of explosives on ambient surfaces by desorption electrospray ionization mass spectrometryChemical Communications, 2005
- Thin-Layer Chromatography and Mass Spectrometry Coupled Using Desorption Electrospray IonizationAnalytical Chemistry, 2005
- Fragmentation of even electron ions: Protonated amines and estersJournal of Mass Spectrometry, 1979