Optimum frequency of office blood pressure measurement using an automated sphygmomanometer
- 1 December 2008
- journal article
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 13 (6), 333-338
- https://doi.org/10.1097/mbp.0b013e3283104247
Abstract
Objective To determine the optimum interval between serial blood pressure measurements using an automated BpTRU sphygmomanometer. Methods Two groups of 200 patients each had automated office measurements taken using the BpTRU device at either 1-min or 2-min intervals from the start of one reading to the start of the next reading with a 24-h ambulatory blood pressure (ABP) recording being performed. Another series of 50 patients had BpTRU readings taken at 1-min and 2-min intervals before and after 24-h ABP monitoring. The difference between the mean awake ABP and the mean automated office BP readings were compared for recordings taken at 1-min versus 2-min intervals. Results In the between-patient comparison (n=400), mean awake ABP was similar to automated BP recordings in the examining room at either 1-min or 2-min intervals except for a slightly lower (−4 mmHg) diastolic BP with the 1-min interval (P<0.01 vs. ABP). In the within-patient comparison (n=50), there was no consistent difference between automated BP readings taken in the examining room at 1-min versus 2-min intervals. Overall, the mean automated BP values tended to be slightly lower than the mean awake ABP. Conclusion Automated measurement of BP in the office setting with devices such as the BpTRU can be taken as frequently as every 1 min without affecting the accuracy of the reading. Small differences in BP between the 1 and 2-min settings and between the automated BpTRU and ABP readings were within accepted clinical standards for validation criteria.Keywords
This publication has 7 references indexed in Scilit:
- Comparison Between an Automated and Manual Sphygmomanometer in a Population SurveyAmerican Journal of Hypertension, 2008
- Performance of the automated BpTRU™ measurement device in the assessment of white-coat hypertension and white-coat effectBlood Pressure Monitoring, 2006
- The BpTRU automatic blood pressure monitor compared to 24 hour ambulatory blood pressure monitoring in the assessment of blood pressure in patients with hypertensionBMC Cardiovascular Disorders, 2005
- Ambulatory Blood Pressure Monitoring for Routine Clinical PracticeHypertension, 2005
- Working Group on Blood Pressure Monitoring of the European Society of Hypertension International Protocol for validation of blood pressure measuring devices in adultsBlood Pressure Monitoring, 2002
- Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitorBlood Pressure Monitoring, 2001
- Statistics Notes: Some examples of regression towards the meanBMJ, 1994