Quality of plasma cholesterol measurements in primary care.
Open Access
- 4 February 1989
- Vol. 298 (6669), 297-298
- https://doi.org/10.1136/bmj.298.6669.297
Abstract
Three surveys were made of the quality of plasma cholesterol measurements performed with a commercial desktop analyser (BCL Reflotron) in primary care. Each survey included three specimens, and results were received from 37, 61, and 69 participants. Although many participants obtained satisfactory results, 8.6% of the results differed by 1.0 mmol/l or more from the target values, and the overall between instrument dispersion of results was 1.3 times that between hospital laboratories. It was found that common sources of error were poor technique and the use of outdated reagent strips. Users of such instruments outside the laboratory need help and advice with training, and guidelines for this are provided. The main recommendations are that users should establish contact with a local clinical chemistry laboratory for training and support and should participate in external quality assessment schemes.Keywords
This publication has 8 references indexed in Scilit:
- How accurately does the Reflotron dry‐chemistry system measure plasma total cholesterol levels when used as a community‐screening device?The Medical Journal of Australia, 1988
- Identification and treatment of risk factors for coronary heart disease in general practice: a possible screening modelBMJ, 1988
- Revised calibration of the Reflotron cholesterol assay evaluated.Clinical Chemistry, 1988
- A MULTICENTER STUDY OF THE NEW REFLOTRON SYSTEM FOR THE MEASUREMENT OF UREA, GLUCOSE, TRIACYLGLYCEROLS, CHOLESTEROL, GAMMA-GLUTAMYLTRANSFERASE AND HEMOGLOBIN1988
- Bias and precision of cholesterol analysis by physician's office analyzers.Clinical Chemistry, 1987
- Strategies for reducing coronary heart disease and desirable limits for blood lipid concentrations: guidelines of the British Hyperlipidaemia Association.BMJ, 1987
- Prevention of cardiovascular disease in general practice: a proposed model.BMJ, 1986
- Quality assessment of blood glucose monitors in use outside the hospital laboratory.Journal of Clinical Pathology, 1983