Continuous intensity map optimization (CIMO): A novel approach to leaf sequencing in step and shoot IMRT
- 13 March 2006
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 33 (4), 859-867
- https://doi.org/10.1118/1.2176057
Abstract
A new leaf-sequencing approach has been developed that is designed to reduce the number of required beam segments for step-and-shoot intensity modulated radiation therapy (IMRT). This approach to leaf sequencing is called continuous-intensity-map-optimization (CIMO). Using a simulated annealing algorithm, CIMO seeks to minimize differences between the optimized and sequenced intensity maps. Two distinguishing features of the CIMO algorithm are (1) CIMO does not require that each optimized intensity map be clustered into discrete levels and (2) CIMO is not rule-based but rather simultaneously optimizes both the aperture shapes and weights. To test the CIMO algorithm, ten IMRT patient cases were selected (four head-and-neck, two pancreas, two prostate, one brain, and one pelvis). For each case, the optimized intensity maps were extracted from the Pinnacle3 treatment planning system. The CIMO algorithm was applied, and the optimized aperture shapes and weights were loaded back into Pinnacle. A final dose calculation was performed using Pinnacle's convolution/superposition based dose calculation. On average, the CIMO algorithm provided a 54% reduction in the number of beam segments as compared with Pinnacle's leaf sequencer. The plans sequenced using the CIMO algorithm also provided improved target dose uniformity and a reduced discrepancy between the optimized and sequenced intensity maps. For ten clinical intensity maps, comparisons were performed between the CIMO algorithm and the power-of-two reduction algorithm of Xia and Verhey [Med. Phys. 25(8), 1424-1434 (1998)]. When the constraints of a Varian Millennium multileaf collimator were applied, the CIMO algorithm resulted in a 26% reduction in the number of segments. For an Elekta multileaf collimator, the CIMO algorithm resulted in a 67% reduction in the number of segments. An average leaf sequencing time of less than one minute per beam was observed.Keywords
This publication has 22 references indexed in Scilit:
- Minimizing beam‐on time in cancer radiation treatment using multileaf collimatorsNetworks, 2004
- Leaf sequencing via difference matrices reduces aperture number and beam-on time in IMRTInternational Journal of Radiation Oncology*Biology*Physics, 2003
- Direct aperture optimization: A turnkey solution for step‐and‐shoot IMRTMedical Physics, 2002
- A quality and efficiency analysis of theIMFAST™ segmentation algorithm in head and neck “step & shoot” IMRT treatmentsMedical Physics, 2002
- Improved leaf sequencing reduces segments or monitor units needed to deliver IMRT using multileaf collimatorsMedical Physics, 2001
- Implementing multiple static field delivery for intensity modulated beamsMedical Physics, 2001
- Leaf sequencing techniques for Mlc-Based IMRTMedical Dosimetry, 2001
- Intensity modulation with the “step and shoot” technique using a commercial mlc: a planning studyInternational Journal of Radiation Oncology*Biology*Physics, 1999
- X-ray field compensation with multileaf collimatorsInternational Journal of Radiation Oncology*Biology*Physics, 1994
- The design and performance characteristics of a multileaf collimatorPhysics in Medicine & Biology, 1994