Optical quality testing of monofocal intraocular lens implants with a 3 mm and a 4 mm aperture
- 1 July 1991
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 17 (4), 485-490
- https://doi.org/10.1016/s0886-3350(13)80856-5
Abstract
The optical quality of 81 monofocal poly(methyl methacrylate) intraocular lenses (IOLs) from eight U.S. firms was tested using a 3 mm aperture, as recommended by the American National Standards Institute Z80.7 standard for IOLs, and a 4 mm aperture. The use of the larger aperture had no effect on the measurements of refractive power and astigmatism. When examined with the 3 mm aperture, the average resolving power of the IOLs was 81% of the diffraction limit; when examined with the 4 mm aperture, the average resolving power was 67% of the diffraction limit. Use of the larger aperture would ensure that a larger area of the IOL had been examined. Incorporating the larger aperture into the ANSI minimum resolution requirement would not appreciably affect the rejection rate for currently manufactured monofocal lenses.Keywords
This publication has 6 references indexed in Scilit:
- Survey of the optical quality of intraocular lens implantsJournal of Cataract & Refractive Surgery, 1991
- Silicone intraocular lens resolution in air and in waterJournal of Cataract & Refractive Surgery, 1988
- Intraocular lens resolution in air and waterJournal of Cataract & Refractive Surgery, 1987
- Optical Aberrations From a Well-Centered Intraocular Lens ImplantAmerican Journal of Ophthalmology, 1986
- Glare Disability In Eyes With Intraocular LensesAmerican Journal of Ophthalmology, 1984
- Apparent Accommodation in Pseudophakic Eyes after Implantation of Posterior Chamber Intraocular LensesAmerican Journal of Ophthalmology, 1983