A computerised dichromatic earpiece densitometer for the measurement of cardiac output
- 1 July 1979
- journal article
- research article
- Published by Oxford University Press (OUP) in Cardiovascular Research
- Vol. 13 (7), 420-426
- https://doi.org/10.1093/cvr/13.7.420
Abstract
This study assesses a precalibrated dichromatic earpiece densitometer and microprocessor for the measurement of cardiac output by indocyanine green dye dilution. The measured cardiac output is compared with values of cardiac output simultaneously determined using a cuvette densitometer. The microprocessor computation of cardiac output agreed very closely with the cardiac output determined by manual calculation from the same dye dilution curves (standard deviation ±1.47%). The reproducibility of the earpiece densitometer (standard deviation ±5.2%) was virtually identical to that of the cuvette densitometer (±5.3%). In a comparison of earpiece and cuvette densitometers for 60 measurements of cardiac output following pulmonary arterial injection of dye and for 50 measurements following femoral venous injection of dye, correlation coefficients were 0.83 and 0.78 and the standard deviations of the differences of simultaneous measurements were 7.2% and 8.3% respectively. The instrument offers an accurate reproducible and relatively noninvasive technique for measuring cardiac output.Keywords
This publication has 6 references indexed in Scilit:
- Measurement of cardiac output by earpiece dye-dilution method with automatic calibration of dye concentration.The Tohoku Journal of Experimental Medicine, 1977
- Use of dichromatic earpiece densitometry for determination of cardiac output.Journal of Applied Physiology, 1967
- A Compensated Dichromatic Densitometer for Indocyanine GreenIRE Transactions on Bio-Medical Electronics, 1962
- Measurement of Cardiac Output in Man by Dye Dilution Curves Using Simultaneous Ear Oximeter and Whole Blood Cuvette TechniquesCirculation Research, 1961
- STUDIES ON THE CIRCULATIONAmerican Journal of Physiology-Legacy Content, 1929
- SIMULTANEOUS DETERMINATION OF THE PULMONARY AND SYSTEMIC CIRCULATION TIMES IN MAN AND OF A FIGURE RELATED TO THE CARDIAC OUTPUTAmerican Journal of Physiology-Legacy Content, 1928