Dosimetry comparison and characterisation of an Al K ultrasoft X-ray beam from an MRC cold-cathode source
- 1 October 1985
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 30 (10), 1029-1041
- https://doi.org/10.1088/0031-9155/30/10/002
Abstract
Ultrasoft X-rays of 0.3-5 keV have provided a unique tool for the investigation of intracellular mechanisms of radiation action in biological organisms, including mammalian cells. However, their use presents unique practical problems in dosimetry and experimental design. Detailed interpretation of the biological results requires reliable dosimetry and well characterised monoenergetic beams. This paper presents a comparison between two fundamentally different dosimetric techniques, namely the ionisation current in an extrapolation chamber and photon counts in a proportional counter. Agreement within 7% was obtained when these two methods were applied to an Al K X-ray beam (1.5 keV) from an MRC cold-cathode transmission target discharge tube as previously used in many biological experiments. Photographic film was calibrated as a relative dosimetric technique and used for investigation of the intensity uniformity of the radiation field. These techniques provide a comprehensive characterisation of the beam in the position of the biological cells, including photon flux (or absorbed dose rate), spectral purity (showing <<1% bremsstrahlung relative to characteristic Al X-rays) and uniformity over the irradiation area (within about 5% for mammalian cell irradiations).This publication has 22 references indexed in Scilit:
- Proton-induced characteristic x-rays: a versatile source of ultrasoft x-rays for biological and biochemical investigationsPhysics in Medicine & Biology, 1984
- Estimation of a single property of low LET radiations which correlates with biological effectivenessPhysics in Medicine & Biology, 1983
- Measurement of W Values of Low-Energy Electrons in Several GasesRadiation Research, 1980
- Technique for producing ultrathin polypropylene filmsReview of Scientific Instruments, 1977
- X-ray calibration of RAR 2490 film for application to laser plasma experimentsApplied Optics, 1977
- Inactivation and Mutation of Cultured Mammalian Cells by Aluminium Characteristic Ultrasoft X-raysInternational Journal of Radiation Biology, 1977
- Inactivation and Mutation of Cultured Mammalian Cells by Aluminium Characteristic Ultrasoft X-raysInternational Journal of Radiation Biology, 1977
- Inactivation and Mutation of Cultured Mammalian Cells by Aluminium Characteristic Ultrasoft X-raysInternational Journal of Radiation Biology, 1977
- Sensitivity of x-ray film. II. Kodak No-Screen film in the 1–100-keV regionJournal of Applied Physics, 1976
- The effect of ionization distribution on chromosome breakage by X-raysJournal of Genetics, 1943