Quantitative Evaluation of Blood Damage in a Centrifugal VAD by Computational Fluid Dynamics
- 1 May 2004
- journal article
- Published by ASME International in Journal of Fluids Engineering
- Vol. 126 (3), 410-418
- https://doi.org/10.1115/1.1758259
Abstract
This study explores a quantitative evaluation of blood damage that occurs in a continuous flow left ventricular assist device (LVAD) due to fluid stress. Computational fluid dynamics (CFD) analysis is used to track the shear stress history of 388 particle streaklines. The accumulation of shear and exposure time is integrated along the streaklines to evaluate the levels of blood trauma. This analysis, which includes viscous and turbulent stresses, provides a statistical estimate of possible damage to cells flowing through the pump. Since experimental data for hemolysis levels in our LVAD are not available, in vitro normalized index of hemolysis values for clinically available ventricular assist devices were compared to our damage indices. This approach allowed for an order of magnitude comparison between our estimations and experimentally measured hemolysis levels, which resulted in a reasonable correlation. This work ultimately demonstrates that CFD is a convenient and effective approach to analyze the Lagrangian behavior of blood in a heart assist device.Keywords
This publication has 20 references indexed in Scilit:
- Computational Fluid Dynamics (CFD) Study of the 4th Generation Prototype of a Continuous Flow Ventricular Assist Device (VAD)Journal of Biomechanical Engineering, 2004
- The Application of Quantitative Oil Streaking to the HeartQuest Left Ventricular Assist DeviceArtificial Organs, 2002
- A Prototype HeartQuest Ventricular Assist Device for Particle Image Velocimetry MeasurementsArtificial Organs, 2002
- Prediction of Hemolysis in Rotary Blood Pumps with Computational Fluid Dynamics AnalysisPublished by Taylor & Francis ,2001
- IN VIVO TESTING OF THE TOTALLY IMPLANTABLE JARVIK 2000® HEART SYSTEMAsaio Journal, 2000
- Computational Flow Optimization of Rotary Blood Pump ComponentsArtificial Organs, 1995
- Three‐Dimensional Numerical Prediction of Stress Loading of Blood Particles in a Centrifugal PumpArtificial Organs, 1995
- Model for a General Mechanical Blood Damage PredictionArtificial Organs, 1995
- Computational Prediction of Hemolysis in a Centrifugal Ventricular Assist DeviceArtificial Organs, 1995
- SOME MECHANICAL EFFECTS THAT INFLUENCE HEMOLYSISAsaio Journal, 1965