Understanding the influence of post‐excite radius and axial confinement on quantitative proteomic measurements using Fourier transform ion cyclotron resonance mass spectrometry
- 28 February 2007
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 21 (7), 1196-1204
- https://doi.org/10.1002/rcm.2957
Abstract
Early studies of Fourier transform ion cyclotron resonance (FT‐ICR) mass spectrometry (MS) explored many of the fundamental issues surrounding the potential of the technique to provide quantitative data. Improvements in instrument technology and the analysis of larger molecules in increasingly complex mixtures warrant not only a revisit to some of these earlier studies, but a more comprehensive examination of the influence of various instrument parameters on quantitative (absolute and relative) measurements in proteomics. We present a detailed examination of the role that acquisition time, excite voltage (i.e. excite radius), trapping voltage, and the type of excitation waveform have on the ability of FT‐ICR to accurately quantify biological molecules. The use of a stable‐isotope‐labeled and unlabeled phenyl isocyanate derivatized peptide allows us to ascribe the effects of FT‐ICR‐MS on quantification, thus eliminating the contribution of ionization differences to ion abundance. To adequately assess the multiple parameters in the large dataset, we develop a multiplicative quality factor that encompasses the total ion abundance, as well as the accuracy and the precision of abundance ratios. This assessment allows facile determination of optimal instrument parameters for quantitative measurements. Copyright © 2007 John Wiley & Sons, Ltd.Keywords
This publication has 45 references indexed in Scilit:
- Multiplexed Protein Quantitation in Saccharomyces cerevisiae Using Amine-reactive Isobaric Tagging ReagentsMolecular & Cellular Proteomics, 2004
- Mass spectrometry-based proteomicsNature, 2003
- Fourier transform ion cyclotron resonance mass spectrometry: A primerMass Spectrometry Reviews, 1998
- Stored waveform inverse Fourier transform (SWIFT) ion excitation in trapped-ion mass spectometry: Theory and applicationsInternational Journal of Mass Spectrometry and Ion Processes, 1996
- Ion traps for Fourier transform ion cyclotron resonance mass spectrometry: principles and design of geometric and electric configurationsInternational Journal of Mass Spectrometry and Ion Processes, 1995
- Optimal cyclotron radius for high resolution FT-ICR spectrometryInternational Journal of Mass Spectrometry and Ion Processes, 1993
- Open trapped ion cell geometries for Fourier transform ion cyclotron resonance mass spectrometryInternational Journal of Mass Spectrometry and Ion Processes, 1992
- Segmented fourier transform and its application to fourier transform ion cyclotron resonance (FT-ICR) mass spectrometry: ion abundances and mass measurementsInternational Journal of Mass Spectrometry and Ion Processes, 1989
- Experimental determination of the effects of space charge on ion cyclotron resonance frequenciesInternational Journal of Mass Spectrometry and Ion Processes, 1983
- Theory of space-charge shift of ion cyclotron resonance frequenciesInternational Journal of Mass Spectrometry and Ion Processes, 1983