The absolute intensity calibration of photographic emulsions in the vacuum ultraviolet
- 1 February 1971
- journal article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 4 (2), 89-92
- https://doi.org/10.1088/0022-3735/4/2/004
Abstract
The response has been measured of Kodak S.W.R. and SC-7 films and Ilford Q1 plates at 584, 736, and 920 AA and of SWR film at 1216 AA. The absolute values of the radiance were measured with a rare gas ionization chamber. The inertia was found to be the most useful measure of emulsion speed.Keywords