Accuracy of the newer generation intraocular lens power calculation formulas in long and short eyes
- 1 March 1991
- journal article
- research article
- Published by Wolters Kluwer Health in Journal of Cataract & Refractive Surgery
- Vol. 17 (2), 187-193
- https://doi.org/10.1016/s0886-3350(13)80249-0
Abstract
The accuracy of two newer generation theoretical intraocular lens (IOL) power calculation formulas and of the empirical SRK I and II formulas was evaluated in a series of 500 IOL implantations including a series of unusually long and short eyes. The prediction error of the theoretical formulas was found to be largely unaffected by the variation in axial length and corneal power, while the prediction of the SRK I formula was less accurate in the short and long eyes. The prediction of the SRK II formula was more accurate than the SRK I in that no systematic offset error with axial length could be demonstrated. However, because of a relatively larger scatter in the long eyes and a significant bias with the corneal power, the absolute error of the SRK II formula was higher than that of the theoretical formulas in the long eyes. The higher accuracy of the newer generation theoretical formulas was attributed to their improved prediction of the pseudophakic anterior chamber depth.Keywords
This publication has 20 references indexed in Scilit:
- Theoretical versus SRK I and SRK II calculation of intraocular lens powerJournal of Cataract & Refractive Surgery, 1990
- Effectiveness of intraocular lens calculation in high ametropiaJournal of Cataract & Refractive Surgery, 1989
- Computerised intraocular lens calculation: clinical results and predictability.British Journal of Ophthalmology, 1989
- Comparison of the SRK II™ formula and other second generation formulasJournal of Cataract & Refractive Surgery, 1988
- A three-part system for refining intraocular lens power calculationsJournal of Cataract & Refractive Surgery, 1988
- Theoretical vs Empirical Prediction of Aphakic RefractionArchives of Ophthalmology (1950), 1987
- Theoretical, computer-assisted prediction versus SRK prediction of postoperative refraction after intraocular lens implantationJournal of Cataract & Refractive Surgery, 1987
- Theoretical approach to intraocular lens calculation using Gaussian opticsJournal of Cataract & Refractive Surgery, 1987
- A New Posterior Chamber Intraocular Lens Formula for Axial MyopesOphthalmology, 1984
- Refraction After Intraocular Lens Implantation: Results with a Computerized System for Ultrasonic Biometry and for Implant Lens Power CalculationPublished by Springer Nature ,1983