Accuracy of pulse oximetry at various haematocrits and during haemolysis in anin vitro model
- 1 March 1993
- journal article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 31 (2), 135-141
- https://doi.org/10.1007/bf02446671
Abstract
In situations in which it may be impossible and/or unethical to evaluate pulse oximetry in humans, an in vitro model with circulating blood may be a necessity. The main objective was to develop such an in vitro model and, in this model, validate the pulse oximetry technique at various haematocrit levels. The pulsating character of arterial blood flow in a tubing system was simulated by using a specially constructed pressure-regulated roller pump. The tubing system was designed to minimise damage to red blood cells. The pulse oximeter readings (SpO2) were compared with oxygen saturation analyses by a haemoximeter (SaO2). The pulse oximetry readings were recorded at various haematocrit levels and during haemolysis in the SaO2 range 60–100 per cent. At a haematocrit level of 41–44 per cent, there was no correlation between SaO2 and SpO2 readings. After diluting the blood with normal saline to a haematocrit of 10–11 per cent, a good correlation between SaO2 and SpO2 was found. Following haemolysis, the agreement between SaO2 and SpO2 was further improved. Using the developed in vitro model, the results indicate that the accuracy of a pulse oximeter may be dependent on the haematocrit level.This publication has 41 references indexed in Scilit:
- The influence of changes in blood flow on the accuracy of pulse oximetry in humansActa Anaesthesiologica Scandinavica, 1992
- In vitro investigation of the factors affecting pulse oximetryJournal of Biomedical Engineering, 1991
- Effect of anemia on pulse oximeter accuracy at low saturationJournal of Clinical Monitoring and Computing, 1990
- Assessment of peripheral blood flow using a pulse oximeterJournal of Clinical Monitoring and Computing, 1990
- An in vitro tissue model for evaluating the effect of carboxyhemoglobin concentration on pulse oximetryIEEE Transactions on Biomedical Engineering, 1989
- Application of AlGaAs/GaAs HBTs to high-speed CML logic family fabricationIEEE Transactions on Electron Devices, 1989
- Changing Perspectives in Monitoring OxygenationAnesthesiology, 1989
- Accuracy of Pulse Oximetry During Arterial Oxyhemoglobin Desaturation in DogsAnesthesiology, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Fabrication and characterization of nonalloyed Cr/Au ohmic contacts to n- and p-type In0.53Ga0.47AsIEEE Transactions on Electron Devices, 1986