Combination of Dynamic and Integral Methods for Generating Reproducible Functional CBF Images

Abstract
A new method to measure regional CBF is presented, applying both dynamic and integral analyses to a dynamic sequence of positron emission tomographic scans collected during and following the administration of H215O (inhalation of C15O2). The dynamic analysis is used to correct continuously monitored arterial whole-blood activity for delay and dispersion relative to tissue scans. An integral analysis including corrections for this delay and dispersion is then used to calculate CBF on a pixel-by-pixel basis. Normal values and reproducibility over a 2-h period are presented, together with the results of validation and simulation studies. The results indicate that the single–tissue compartment model adequately describes the distribution of H215O in the brain, without recourse to postulating a nonexchanging water pool.

This publication has 19 references indexed in Scilit: