Flexible Polymer Sensors for In Vivo Intravascular Shear Stress Analysis
- 3 September 2008
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in Journal of Microelectromechanical Systems
- Vol. 17 (5), 1178-1186
- https://doi.org/10.1109/jmems.2008.927749
Abstract
Hemodynamic forces, specifically fluid shear stress, play an important role in the focal nature of arterial plaque formation known as atherosclerosis. We hereby developed biocompatible and flexible intravascular microelectromechanical systems sensor to measure real-time shear stress in the aortas of New Zealand white (NZW) rabbits. Titanium (Ti) and platinum (Pt) were deposited on silicon wafers and patterned to form the sensing elements. The polymer, parylene C, provided insulation to the electrode leads and flexibility to the sensors. Based on heat transfer principle, the heat dissipation from the sensors to the blood flow altered the resistance of the sensing elements, from which shear stress was calibrated. The resistance of the sensing element was measured at approximately 1.0 kOmega , and the temperature coefficient of resistance was at approximately 0.16%/degC. The individual sensors were packaged to the catheter for intravascular deployment in the aortas of NZW rabbits (n = 5) . The sensor was capable of resolving spatial- and time-varying components of shear stress in the abdominal aorta. Computational fluid dynamic code based on non-Newtonian fluid properties showed comparable results within an acceptable range of experimental errors ( plusmn9%) for the maximal and minimal values in shear stress during one cardiac cycle. Therefore, we demonstrated the capability of biocompatible sensors for real-time shear stress measurement in vivo with a potential to advance the understanding between the blood flow and vascular disease.Keywords
This publication has 21 references indexed in Scilit:
- Spatial Variations in Shear Stress in a 3-D Bifurcation Model at Low Reynolds NumbersAnnals of Biomedical Engineering, 2005
- Experiments and simulations of MEMS thermal sensors for wall shear-stress measurements in aerodynamic control applicationsJournal of Micromechanics and Microengineering, 2004
- A parylene MEMS flow sensing arrayPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2004
- Micro Sensors: Linking Real-Time Oscillatory Shear Stress with Vascular Inflammatory ResponsesAnnals of Biomedical Engineering, 2004
- IC-integrated flexible shear-stress sensor skinJournal of Microelectromechanical Systems, 2003
- Relation between near-wall residence times of monocytes and early lesion growth in the rabbit aorto-celiac junction.Annals of Biomedical Engineering, 2003
- MICRO-ELECTRO-MECHANICAL-SYSTEMS (MEMS) AND FLUID FLOWSAnnual Review of Fluid Mechanics, 1998
- BLOOD FLOW IN ARTERIESAnnual Review of Fluid Mechanics, 1997
- Identification of vascular endothelial genes differentially responsive to fluid mechanical stimuli: cyclooxygenase-2, manganese superoxide dismutase, and endothelial cell nitric oxide synthase are selectively up-regulated by steady laminar shear stress.Proceedings of the National Academy of Sciences, 1996