Investigation of Noninvasive in vivo Blood Hematocrit Measurement Using NIR Reflectance Spectroscopy and Partial Least-Squares Regression
- 1 February 2000
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 54 (2), 294-299
- https://doi.org/10.1366/0003702001949258
Abstract
Hematocrit (Hct), the volume percent of red cells in blood, is monitored routinely for blood donors, surgical patients, and trauma victims and requires blood to be removed from the patient. An accurate, noninvasive method for directly measuring hematocrit on patients is desired for these applications. The feasibility of noninvasive hematocrit measurement by using near-infrared (NIR) spectroscopy and partial least-squares (PLS) techniques was investigated, and methods of in vivo calibration were examined. Twenty Caucasian patients undergoing cardiac surgery on cardiopulmonary bypass were randomly selected to form two study groups. A fiber-optic probe was attached to the patient's forearm, and NIR spectra were continuously collected during surgery. Blood samples were simultaneously collected and reference Hct measurements were made with the spun capillary method. PLS multivariate calibration techniques were applied to investigate the relationship between spectral and Hct changes. Single patient calibration models were developed with good cross-validated estimation of accuracy (∼ 1 Hct%) and trending capability for most patients. Time-dependent system drift, patient temperature, and venous oxygen saturation were not correlated with the hematocrit measurements. Multi-subject models were developed for prediction of independent subjects. These models demonstrated a significant patient-specific offset that was shown to be partially related to spectrometer drift. The remaining offset is attributed to the large spectral variability of patient tissue, and a significantly larger set of patients would be required to adequately model this variability. After the removal of the offset, the cross-validated estimation of accuracy is 2 Hct%.Keywords
This publication has 12 references indexed in Scilit:
- Measurement of pH in Whole Blood by Near-Infrared SpectroscopyApplied Spectroscopy, 1999
- Phantom Glucose Calibration Models from Simulated Noninvasive Human Near-Infrared SpectraAnalytical Chemistry, 1998
- Partial Least-Squares Modeling of Near-Infrared Reflectance Data for Noninvasive in Vivo Determination of Deep-Tissue pHApplied Spectroscopy, 1998
- Feasibility of non-invasive measurement of tissue pH using near-infrared reflectance spectroscopyJournal of Clinical Monitoring and Computing, 1996
- Primary Sjögren’s syndrome, ulcerative colitis and selective IgA deficiencyPublished by Oxford University Press (OUP) ,1996
- Haematocrit measurements during cardiopulmonary bypass surgery: comparison of three stat methods with a blood cell counterPerfusion, 1994
- Reagentless Near-Infrared Determination of Glucose in Whole Blood Using Multivariate CalibrationApplied Spectroscopy, 1992
- Noninvasive Glucose Monitoring in Diabetic Patients: A Preliminary EvaluationClinical Chemistry, 1992
- Effects of Temperature on Optical Absorbance Spectra of Oxy-, Carboxy-, and DeoxyhemoglobinClinical Chemistry, 1992
- Angiotensin decreases cyclic GMP accumulation produced by atrial natriuretic factorAmerican Journal of Physiology-Cell Physiology, 1987