How much light from rectangular scintillation counters? (for whole body counting)
- 1 March 1975
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 20 (2), 282-295
- https://doi.org/10.1088/0031-9155/20/2/010
Abstract
Theoretical calculations have been made of the efficiency and uniformity of light collection at one end of a large rectangular liquid scintillation counter (80x20x18 cm). Three different arrangements were considered; (1) a system in which all surfaces of the counter other than the collecting surface were transparent to permit total internal reflection, (2) a similar system with a mirror placed close to the end opposite the collecting surface and (3) a system in which external reflectors were placed close to surfaces other than the collecting surface. In all cases, the effects of attenuation of light in the counter medium, inefficient total internal reflection and the reflectivities of external reflectors were considered and illustrated for selected values of these parameters. In an experimental counter of the same dimensions, the amplitude of the spectrum of a collimated 42-K source was used to measure the relative efficiencies of light collection for various reflector arrangements. In general the experimental values agreed with theory and the non-uniformity of response was within the predicted range. It is estimated that about 30% of scintillation light can be collected at one end of such a counter, with a non-uniformity not greater than 10%, if magnesium oxide is used as the external reflector at all other surfaces.Keywords
This publication has 12 references indexed in Scilit:
- Angular distribution of “trapped” light in rectangular scintillation countersNuclear Instruments and Methods, 1971
- Performance of a high sensitivity liquid scintillation whole-body counterJournal of Physics E: Scientific Instruments, 1968
- On the uniformity of response of large plastic scintillation counters to minimum ionizing particlesNuclear Instruments and Methods, 1967
- Optimum design of thin large area scintillation countersNuclear Instruments and Methods, 1966
- Liquid scintillators based on medicinal paraffin for use in whole-body countersJournal of Scientific Instruments, 1966
- Efficiency and transparency of cheap liquid scintillatorsJournal of Scientific Instruments, 1964
- Performance of large liquid scintillation countersNuclear Instruments and Methods, 1963
- Performance of Large Area Scintillation CountersProceedings of the Physical Society, 1960
- A LARGE-AREA LIQUID SCINTILLATION COUNTER AND SOME MEASUREMENTS ON HIGH-ENERGY COSMIC-RAY PARTICLESCanadian Journal of Physics, 1958
- The Trapping of Fluorescent Light Produced within Objects of High Geometrical SymmetryJournal of the Optical Society of America, 1949