Noninvasive Measurement of Myocardial Activity Concentrations and Perfusion Defect Sizes in Rats With a New Small-Animal Positron Emission Tomograph

Abstract
Background— We explored the feasibility of measuring regional tracer activity concentrations and flow defects in myocardium of rats with a high spatial resolution small-animal PET system (microPET). Methods and Results— Myocardial images were obtained after intravenous 18F-fluorodeoxyglucose (18FDG) in 11 normal rats (group 1) and assembled into polar maps. Regional 18F activity concentrations were measured in 9 regions of interest and compared with tissue activity concentrations measured by well counting. In another 9 rats (group 2), myocardial perfusion images were acquired with 13N-ammonia at baseline and during coronary occlusion. On the polar maps recorded during coronary occlusion, the size of perfusion defects was measured as the myocardium with <50% of maximum activity and expressed as percent total myocardium and was correlated with the area at risk defined by postmortem staining. The diagnostic quality of 18FDG and 13N-ammonia microPET images was good to excellent; the images were easily assembl...

This publication has 14 references indexed in Scilit: