Biodegradable Interference Screw Fixation Exhibits Pull-Out Force and Stiffness Similar to Titanium Screws
- 1 January 1998
- journal article
- research article
- Published by SAGE Publications in The American Journal of Sports Medicine
- Vol. 26 (1), 119-128
- https://doi.org/10.1177/03635465980260010401
Abstract
Recently, increased interest in biodegradable interference screws for bone-tendon-bone graft fixation has led to numerous screws becoming available. The implants are made from different polymers and have different designs, which might influence their mechanical properties. Several studies have reported a wide range of mechanical results for these screws using different biomechanical models. The aim of the present study is to compare reliable biomechanical data for six different biodegradable interference screws, consisting of five different polymers, with a conventional titanium screw in a standardized model. Seventy proximal calf tibias were used to determine maximal pull-out force, stiffness of fixation, and insertion torque for interference screw fixation of bone-tendon-bone grafts. Additionally, maximal torque at failure was determined. Data were analyzed with respect to aspects of screw design, such as drive and thread shape. Five of the six biodegradable screws provided initial pull-out force and stiffness of fixation comparable with that of a conventional titanium screw. Torque at failure can be greatly increased by adapting the drive design to the mechanical properties of the polymeric raw material. A correlation between pull-out force and thread height indicates that fixation rigidity depends on screw design, even in a biodegradable implant.Keywords
This publication has 42 references indexed in Scilit:
- Biomechanical comparison between BioScrew and titanium alloy interference screws for bone—patellar tendon—bone graft fixation in anterior cruciate ligament reconstructionArthroscopy: The Journal of Arthroscopic & Related Surgery, 1997
- Fixation strength of interference screw fixation in bovine, young human, and elderly human cadaver knees: Influence of insertion torque, tunnel-bone block gap, and interferenceKnee Surgery, Sports Traumatology, Arthroscopy, 1996
- Preliminary results of an absorbable interference screwArthroscopy: The Journal of Arthroscopic & Related Surgery, 1995
- The deadman theory of suture anchors: observations along a South Texas fence lineArthroscopy: The Journal of Arthroscopic & Related Surgery, 1995
- Late degradation tissue response to poly(?-lactide) bone plates and screwsBiomaterials, 1995
- Optimal graft fixation—The effect of gap size and screw size on bone plug fixation in ACL reconstructionArthroscopy: The Journal of Arthroscopic & Related Surgery, 1994
- The biomechanics of interference screw fixation of patellar tendon anterior cruciate ligament graftsThe American Journal of Sports Medicine, 1993
- Optimizing Bone Screw Pullout ForceJournal of Orthopaedic Trauma, 1990
- Mechanical properties of biodegradable polymers and composites proposed for internal fixation of boneJournal of Applied Biomaterials, 1990
- A study of some factors which affect the strength of screws and their insertion and holding power in boneJournal of Biomechanics, 1968