Electron contamination from the lead cutout used in kilovoltage radiotherapy
- 1 December 1999
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 45 (1), 1-8
- https://doi.org/10.1088/0031-9155/45/1/301
Abstract
In kilovoltage radiotherapy, low-energy electrons are generated on the lead cutout plate frequently used to shape the x-ray field. They can cause surface dose enhancement and affect the percentage depth dose (PDD) and cutout factor when these parameters are measured with a parallel-plate chamber. Although many previous studies concerning a similar surface dose enhancement have suggested using a thin electron filter to eliminate the effect of these low-energy electrons, no details have been given regarding the adequacy of the filter. This paper describes the procedures required to test an electron filter for our Darpec 2000 superficial x-ray machine. A piece of plain paper was found to be an adequate electron filter, which could block all the low-energy electrons without affecting the PDD and cutout factor measurement. The surface dose enhancement caused by the presence of the cutout plate was studied with such a filter. The enhancement was found to increase with x-ray energy and decrease with cutout size as well as applicator size. At the highest x-ray energy in this study (120 kV, HVL 8.6 mmAl), the enhancement ranges from less than 1% for a 8 cm diameter cutout to a maximum of 14% for a 2 cm diameter cutout with a 10 cm applicator. For a 5 cm diameter applicator it ranges from about 1% to about 11%. The difference between the measured and predicted cutout factor displays similar characteristics. Since the low-energy electrons do not generate any useful treatment dose but their presence can lead to potential dosimetry problems, we suggest that a filter tested with the procedure described in this paper should be used when percentage depth dose and cutout factor for kilovoltage x-ray are measured with a parallel-plate chamber.Keywords
This publication has 8 references indexed in Scilit:
- Broad beam attenuation of kilovoltage photon beams: effect of ion chambers.The British Journal of Radiology, 1998
- Measurement of percentage depth dose and lateral beam profile for kilovoltage x-ray therapy beamsPhysics in Medicine & Biology, 1997
- Forward dose perturbation at high atomic number interfaces in kilovoltage x-ray beamsMedical Physics, 1997
- The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al - 4 mm Cu HVL; 10 - 300 kV generating potential)Physics in Medicine & Biology, 1996
- Considerations for superficial photon dosimetryMedical Physics, 1995
- Complications in low energy X-ray dosimetry caused by electron contaminationPhysics in Medicine & Biology, 1991
- Surface dose in intracavitary orthovoltage radiotherapyMedical Physics, 1990
- New Values for Epidermal Thickness and Their ImportanceHealth Physics, 1973