The Observation of Thick Specimens by High Voltage Electron Microscopy. II. Experiment with Molybdenite Films at 50-1200 kV.
- 1 March 1968
- journal article
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 7 (3)
- https://doi.org/10.1143/jjap.7.200
Abstract
The maximum usable thickness of specimens, below which dislocation images were observable by transmission, was studied experimentally in the region 50–500 kV in the previous work [Japan J. appl. Phys. 6 (1967), 557]. The present work is a continuation of the previous one. The maximum usable thickness was determined by comparing the quality of images. The experimental result turned out to be 1.5, 2.5, 4.0, 5.0 and 6.0 µm at 100, 200, 500, 1000 and 1200 kV, respectively. These values are approximately proportional to β2=(v/c)2, although the value at 1200 kV is a little too high. The factors limiting the observation of thick specimens, i.e. image intensity, resolution and contrast, are discussed. It is not yet clear whether the above result is common to other materials or is valid in the higher voltage region.Keywords
This publication has 4 references indexed in Scilit:
- The Observation of Thick Specimens by High Voltage Electron Microscopy. Experiment with Molybdenite Films at 50-500 kVJapanese Journal of Applied Physics, 1967
- Metallurgical Investigations with a 500 kV Electron MicroscopeJapanese Journal of Applied Physics, 1967
- Energy Dependence of Extinction Distance and Transmissive Power for Electron Waves in CrystalsJournal of Applied Physics, 1964
- Excess and Defect Kikuchi-bandsJournal of the Physics Society Japan, 1953