4D-CT imaging of a volume influenced by respiratory motion on multi-slice CT
Top Cited Papers
- 26 January 2004
- journal article
- Published by Wiley in Medical Physics
- Vol. 31 (2), 333-340
- https://doi.org/10.1118/1.1639993
Abstract
We propose a new scanning protocol for generating 4D-CT image data sets influenced by respiratory motion. A cine scanning protocol is used during data acquisition, and two registration methods are used to sort images into temporal phases. A volume is imaged in multiple acquisitions of 1 or 2 cm length along the cranial-caudal direction. In each acquisition, the scans are continuously acquired for a time interval greater than or equal to the average respiratory cycle plus the duration of the data for an image reconstruction. The x ray is turned off during CT table translation and the acquisition is repeated until the prescribed volume is completely scanned. The scanning for 20 cm coverage takes about 1 min with an eight-slice CT or 2 mins with a four-slice CT. After data acquisition, the CT data are registered into respiratory phases based on either an internal anatomical match or an external respiratory signal. The internal approach registers the data according to correlation of anatomy in the CT images between two adjacent locations in consecutive respiratory cycles. We have demonstrated the technique with ROIs placed in the region of diaphragm. The external approach registers the image data according to an externally recorded respiratory signal generated by the Real-Time Position Management (RPM) Respiratory Gating System (Varian Medical Systems, Palo Alto, CA). Compared with previously reported prospective or retrospective imaging of the respiratory motion with a single-slice or multi-slice CT, the 4D-CT method proposed here provides (1) a shorter scan time of three to six times faster than the single-slice CT with prospective gating; (2) a shorter scan time of two to four times improvement over a previously reported multi-slice CT implementation, and (3) images over all phases of a breathing cycle. We have applied the scanning and registration methods on phantom, animal and patients, and initial results suggest the applicability of both the scanning and the registration methods.Keywords
This publication has 11 references indexed in Scilit:
- A method for the reconstruction of four‐dimensional synchronized CT scans acquired during free breathingMedical Physics, 2003
- Respiration‐correlated spiral CT: A method of measuring respiratory‐induced anatomic motion for radiation treatment planningMedical Physics, 2002
- Acquiring a four-dimensional computed tomography dataset using an external respiratory signalPhysics in Medicine & Biology, 2002
- Potential radiotherapy improvements with respiratory gatingAustralasian Physics & Engineering Sciences in Medicine, 2002
- Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancerMedical Physics, 2002
- Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imagingInternational Journal of Radiation Oncology*Biology*Physics, 2002
- Heart Rate Adaptive Optimization of Spatial and Temporal Resolution for Electrocardiogram-Gated Multislice Spiral CT of the HeartJournal of Computer Assisted Tomography, 2001
- Cardiac Imaging by Means of Electrocardiographically Gated Multisection Spiral CT: Initial ExperienceRadiology, 2000
- Electrocardiogram‐correlated image reconstruction from subsecond spiral computed tomography scans of the heartMedical Physics, 1998
- Optimal short scan convolution reconstruction for fan beam CTMedical Physics, 1982