Comparison and quantitation of wear debris of failed total hip and total knee arthroplasty
- 1 June 1996
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 31 (2), 257-263
- https://doi.org/10.1002/(sici)1097-4636(199606)31:2<257::aid-jbm13>3.0.co;2-i
Abstract
To compare the physical properties of debris particles associated with failed total hip and total knee arthroplasty, we applied a recently developed assay to electronically characterize the size, number, and composition of debris particles isolated from tissues adjacent to failed implants. We identified 21 samples (from 20 patients) of hip synovia and 35 samples (from 32 patients) of knee tissues that had been obtained at the time of revision arthroplasty. There were 12 females and 9 males in the hip group, and 16 females and 19 males in the knee group. Primary arthroplasty was performed for osteoarthritis (OA, 15 cases) or rheumatoid arthritis (RA, 6 cases) in the hip, and for OA (23) or RA (12) in the knee. Patients ranged in age from 23 to 85 (mean 59 years) for total hip, and from 27 to 84 (mean 61 years) for total knee arthroplasty. Implantation duration was from 5 to 123 months (mean 37.8) for total hip, and from 11 to 123 months (mean 63.1) for total knee arthroplasty. All of the implants were composed of cobalt-chromium alloy articulating with ultrahigh-molecular-weight polyethylene. The number of particles smaller than 10 μm ranged from 1.04 × 108/g to 1.91 × 1010/g in the hip, and from 6.69 × 108/g to 2.13 × 1010/g in the knee. Energy-dispersive X-ray spectroscopy and polarized light analysis showed both polyethylene and metal particles in most cases. The mean diameter of particles smaller than 10 μm was 0.72 ± 0.2 μm in the hip, and 0.74 ± 0.1 μm in the knee. Evaluation of particles larger than 10 μm showed a larger range of particle size in knee tissues (maximum 6.1 mm, mean 283 μm), than in the hip tissues (maximum 826 μm, mean 81 μm) (p <0.001). Very small particles are common in both groups, but it appears that a larger range of particle sizes is present adjacent to failed knee than to failed hip prostheses. The higher frequency of large particles in failed knee prostheses probably reflects the perceived higher rate of delamination and fragmentation of tibial and patellar compared to that of acetabular polyethylene. © 1996 John Wiley & Sons, Inc.Keywords
This publication has 29 references indexed in Scilit:
- Size and shape of biomaterial wear debrisClinical Materials, 1994
- Polyethylene wear debris and tissue reactions in knee as compared to hip replacement prosthesesJournal of Applied Biomaterials, 1994
- Evaluation of carcinogenicity and chronic toxicity associated with orthopedic implants in miceJournal of Biomedical Materials Research, 1994
- Tissue digestion for wear debris particle isolationJournal of Biomedical Materials Research, 1994
- Macrophage/particle interactions: Effect of size, composition and surface areaJournal of Biomedical Materials Research, 1994
- Metal ion concentrations in retrieved polyethylene total hip inserts and implications for artifactually high readings in tissueJournal of Biomedical Materials Research, 1993
- A method for production and characterization of metal prosthesis wear particlesJournal of Orthopaedic Research, 1993
- Cobalt-Alloy Metal Debris in Periarticular Tissues From Total Hip Revision ArthroplastiesPublished by Wolters Kluwer Health ,1992
- Wear of Polyethylene in the Total Hip ProsthesisPublished by Wolters Kluwer Health ,1982
- The production and biology of polyethylene wear debrisArchives of orthopaedic and trauma surgery, 1978