Effect of unrestricted activity on accuracy of ambulatory blood pressure measurement.
- 1 November 1991
- journal article
- research article
- Published by Wolters Kluwer Health in Hypertension
- Vol. 18 (5), 593-597
- https://doi.org/10.1161/01.hyp.18.5.593
Abstract
A validation study of the Takeda TM-2420 ambulatory blood pressure recorder was performed on 10 subjects using the Oxford ambulatory intra-arterial recording apparatus during unrestricted activity. Electronic linkage of the two recorders ensured simultaneous blood pressure readings, taken from opposite arms. Although there was close approximation of intra-arterial and automated sphygmomanometric recordings over the range of blood pressure encountered in this study, there was a wide scatter of points and a tendency for the machine to underestimate systolic pressure by more than 15 mm Hg in the hypertensive range (systolic blood pressure more than 160 mm Hg) was detected. These findings suggest that automated recording of blood pressure during unrestricted activity may have a proportion of artifactual readings. Although simultaneous intra-arterial blood pressure recording may not be appropriate for widespread use in device validation, this study illustrated some potential disadvantages of the current validation recommendations, namely, the absence of assessment of device accuracy during unrestricted and ambulatory activity.Keywords
This publication has 7 references indexed in Scilit:
- Bench and ambulatory field evaluation of the A & D TM-2420 automated sphygmomanometerJournal Of Hypertension, 1990
- The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systemsJournal Of Hypertension, 1990
- The assessment of portable, non-invasive blood pressure recordersJournal Of Hypertension, 1990
- Assessment of four ambulatory blood pressure monitors and measurements by clinicians versus intraarterial blood pressure at rest and during exerciseThe American Journal of Cardiology, 1990
- 24-Hour blood pressure monitoringJournal Of Hypertension, 1988
- Indirect measurement of blood pressure during exercise testing can be misleading.Heart, 1985
- Evaluation of the Remler M2000 blood pressure recorder. Comparison with intraarterial blood pressure recordings both at hospital and at home.Hypertension, 1984