A micropower dry-electrode ECG preamplifier
- 1 January 2000
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 47 (2), 155-162
- https://doi.org/10.1109/10.821734
Abstract
This paper describes the development of a very low-power preamplifier intended for use in pasteless-electrode recording of the human electrocardiogram. The expected input signal range is 100 /spl mu/V-10 mV from a lead-II electrode configuration. The amplifier provides a gain of 43 dB in a 3-dB bandwidth of 0.05 Hz-2 kHz with a defined high input impedance of 75 M/spl Omega/. It uses a driven common electrode to enhance rejection of common-mode interfering signals, including low-frequency motion artifact, achieving a common-mode rejection ratio (CMRR) of better than 80 dB over its entire bandwidth. The gain and phase characteristics meet the recommendations of the American Heart Association, ensuring low distortion of the output ECG signal and making it suitable for clinical monitoring. The amplifier has a power consumption of 30 /spl mu/W operating from a 3.3-V battery and is intended for use in small, lightweight, portable electrocardiographic equipment and heart-rate monitoring instrumentation.Keywords
This publication has 17 references indexed in Scilit:
- Non-polarisable dry electrode based on NASICON ceramicMedical & Biological Engineering & Computing, 1995
- Common mode rejection ratio in differential amplifiersIEEE Transactions on Instrumentation and Measurement, 1991
- Driven-right-leg circuit designIEEE Transactions on Biomedical Engineering, 1983
- Reduction of Interference Due to Common Mode Voltage in Biopotential AmplifiersIEEE Transactions on Biomedical Engineering, 1983
- Ground-Free ECG Recording with Two ElectrodesIEEE Transactions on Biomedical Engineering, 1980
- A model of bioelectrode motion artefact and reduction of artefact by amplifier input stage designJournal of Biomedical Engineering, 1979
- Minimizing Electrode Motion Artifact by Skin AbrasionIEEE Transactions on Biomedical Engineering, 1977
- 60-Hz Interference in ElectrocardiographyIEEE Transactions on Biomedical Engineering, 1973
- Insulated electrocardiogram electrodesMedical & Biological Engineering & Computing, 1972
- The low-frequency response of electrocardiographs, a frequent source of recording errorsAmerican Heart Journal, 1966