The Status of High-Energy Photon and Electron Beam Dosimetry Five Years After the Implementation of the Iaea Code of Practice in the Nordic Countries
- 1 January 1993
- journal article
- research article
- Published by Taylor & Francis in Acta Oncologica
- Vol. 32 (5), 483-500
- https://doi.org/10.3109/02841869309096107
Abstract
The status of the dosimetry of high-energy photon and electron beams is analysed, taking into account the main developments in the field since the implementation of the IAEA Code of Practice in the Nordic countries. In electron beam dosimetry, energy-range relationships are discussed; Monte-Carlo results with different codes are compared with the experimentally derived empirical expression used in most protocols. Updated calculations of water-to-air stopping-power ratios following the changes in the Monte-Carlo code used to compute actual sw,air values are compared with the data included in most dosimetry protocols. The validity of the commonly used procedure to select stopping-power ratios for a clinical beam from the mean energy at the phantom surface and the depth of measurement, is analysed for ‘realistic’ electron beams. In photon beam dosimetry, calculated correction factors including the effect of the wall plus waterproofing sleeve and existing data on the shift of the effective point of measurement of an ionization chamber, are discussed. New calculations of medium-to-air stopping-power ratios and their correlation with the quality of the beam obtained from the convolution of Monte-Carlo kernels are presented together with their possible practical implications in dosimetry. Trends in Primary Standard Dosimetry Laboratories towards implementing calibrations in terms of absorbed dose to water are presented, emphasising controversial proposals for the specification of photon beam qualities. Plane-parallel ionization chambers are discussed regarding aspects that affect determinations of absorbed dose, either through the different methods used for the calibration of these chambers or by means of correction factors. Recent studies on the effect of the central electrode in Farmer-type cylindrical chambers are described.Keywords
This publication has 69 references indexed in Scilit:
- Perturbation correction factors for some ionization chambers commonly applied in electron beamsPhysics in Medicine & Biology, 1991
- A comparison of ionization‐chamber and water‐calorimeter dosimetry for high‐energy x raysMedical Physics, 1991
- Comparison of IAEA 1987 and AAPM 1983 protocols for dosimetry calibration of radiotherapy beamsMedical Physics, 1991
- Influence of energy and angular spread on stopping-power ratios for electron beamsPhysics in Medicine & Biology, 1989
- Experimental verification of the air kerma to absorbed doseconversion factor Cw,uRadiotherapy and Oncology, 1987
- Comparison of different protocols for the dosimetry of high-energy photon and electron beamsRadiotherapy and Oncology, 1985
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983
- Electron Beams with Mean Energies at the Phantom Surface below 15 MeV Supplement to the Recommendations by the Nordic Association of Clinical Physics (NACP) 1980Acta Radiologica: Oncology, 1981
- Procedures in External Radiation Therapy Dosimetry with Electron and Photon Beams with Maximum Energies Between 1 and 50 MeV Recommendations by the Nordic Association of Clinical Physics (NACP)Acta Radiologica: Oncology, 1980
- Specification of electron beam quality from the central‐axis depth absorbed‐dose distributionMedical Physics, 1976