Real-Time Intravascular Shear Stress in the Rabbit Abdominal Aorta
- 10 June 2009
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 56 (6), 1755-1764
- https://doi.org/10.1109/tbme.2009.2013455
Abstract
Fluid shear stress is intimately linked with the biological activities of vascular cells. A flexible microelectromechanical system (MEMS) sensor was developed to assess spatial- and temporal-varying components of intravascular shear stress (ISS) in the abdominal aorta of adult New Zealand white (NZW) rabbits. Real-time ISS (ISS real-time ) was analyzed in comparison with computational fluid dynamics (CFD) simulations for wall shear stress (WSS). Three-dimensional abdominal arterial geometry and mesh were created using the GAMBIT software. Simulation of arterial flow profiles was established by FLUENT. The Navier-Stokes equations were solved for non-Newtonian blood flow. The coaxial-wire-based MEMS sensor was deployed into the abdominal arteries of rabbits via a femoral artery cutdown. Based on the CFD analysis, the entrance length of the sensor on the coaxial wire (0.4 mm in diameter) was less than 10 mm. Three-dimensional fluoroscope and contrast dye allowed for visualization of the positions of the sensor and ratios of vessel to coaxial wire diameters. Doppler ultrasound provided the velocity profiles for the CFD boundary conditions. If the coaxial wire were positioned at the center of vessel, the CFD analysis revealed a mean ISS value of 31.1 with a systolic peak at 102.8 dyn ldr cm -2 . The mean WSS was computed to be 10.1 dyn ldr cm -2 with a systolic peak at 33.2 dynldrcm -2 , and the introduction of coaxial wire increased the mean WSS by 5.4 dyn ldr cm -2 and systolic peak by 18.0 dyn ldr cm -2 . Experimentally, the mean ISS was 11.9 dyn ldr cm -2 with a systolic peak at 47.0 dyn ldr cm -2 . The waveform of experimental ISS was similar to that of CFD solution with a 30.2% difference in mean and 8.9% in peak systolic shear stress. Despite the difference between CD and experimental results, the flexible coaxial-wire-based MEMS sensors provided a possibility to assess real-time ISS in the abdominal aorta of NZW rabbits.Keywords
This publication has 41 references indexed in Scilit:
- Flexible Polymer Sensors for In Vivo Intravascular Shear Stress AnalysisJournal of Microelectromechanical Systems, 2008
- Recent Developments in Polymer MEMSAdvanced Materials, 2007
- Hemodynamics influences vascular peroxynitrite formation: Implication for low-density lipoprotein apo-B-100 nitrationFree Radical Biology & Medicine, 2006
- 17β-Estradiol reverses shear-stress-mediated low density lipoprotein modificationsFree Radical Biology & Medicine, 2006
- Spatial Variations in Shear Stress in a 3-D Bifurcation Model at Low Reynolds NumbersAnnals of Biomedical Engineering, 2005
- Quantitative Evaluation of Blood Damage in a Centrifugal VAD by Computational Fluid DynamicsJournal of Fluids Engineering, 2004
- A micromachined flow shear-stress sensor based on thermal transfer principlesJournal of Microelectromechanical Systems, 1999
- Three-Dimensional Simulation of Blood Flow in an Abdominal Aortic Aneurysm—Steady and Unsteady Flow CasesJournal of Biomechanical Engineering, 1994
- Fluid shear stress in prosthetic heart valvesJournal of Biomechanics, 1977
- Developing pressure profiles for non‐Newtonian flow in an annular ductAIChE Journal, 1974