Reliability and Validity of the Computer Science and Applications Accelerometer in a Mechanical Setting
- 1 June 2003
- journal article
- Published by Taylor & Francis in Measurement in Physical Education and Exercise Science
- Vol. 7 (2), 101-119
- https://doi.org/10.1207/s15327841mpee0702_4
Abstract
Accelerometry has been proposed as an objective method for measurement of physical activity. Intra- and inter-instrument reliability and validity of the Computer Science and Applications (CSA) Model 7164 accelerometer, along with the nature of the analogue filter in the device, was studied with sinusoidal movements in a mechanical setting. Six CSA units were tested with 17 different frequencies (0.5 - 4 Hz) on 3 radius settings (0.022 - 0.049 m), yielding 51 different acceleration settings (0.1 - 19.7 m· sec-2). Intra-instrument reliability was relatively good (mean CV of 4.4%), although questionable on extreme values of acceleration (less than 1 m · sec-2 and higher than 16 m · sec-2). Analyses on inter-instrument reliability revealed both overall systematic bias and acceleration-specific differences between units, approaching unit error magnitudes of 20% from the mean of all 6 units. The correlation between CSA output and acceleration (i.e., validity) was significant but the relation was clearly nonline...Keywords
This publication has 14 references indexed in Scilit:
- Technical reliability of the CSA activity monitor: The EarlyBird StudyMedicine & Science in Sports & Exercise, 2002
- Assessment of Physical Activity with the Computer Science and Applications, Inc., Accelerometer: Laboratory versus Field ValidationResearch Quarterly for Exercise and Sport, 2000
- Using the Computer Science and Applications (CSA) Activity Monitor in Preschool ChildrenPediatric Exercise Science, 1999
- Piezoelectric properties of zirconium-doped barium titanate single crystals grown by templated grain growthJournal of Applied Physics, 1999
- Calibration of the Computer Science and Applications, Inc. accelerometerMedicine & Science in Sports & Exercise, 1998
- Statistical Methods For Assessing Measurement Error (Reliability) in Variables Relevant to Sports MedicineSports Medicine, 1998
- A critical discussion of intraclass correlation coefficientsStatistics in Medicine, 1994
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Low frequency self-generated vibration during ambulation in normal menJournal of Biomechanics, 1982
- The sources of external work in level walking and running.The Journal of Physiology, 1976